Phase |
Range |
± 180.00? or 0.00? to +360.00? |
Accuracy (23? ± 5?C; 1-year cal interval) |
0.1Hz to 99Hz |
± 0.02? ± 1 digit |
100Hz to 999Hz |
± 0.03? ± 0.005? /kHz ± 1 digit |
1kHz to 99Hz |
± 0.04? ± 0.005? /kHz ± 1 digit |
100kHz to 700kHz |
± 0.25? ± 0.003? /kHz ± 1 digit |
Resolution |
0.01? (0.001? via interfaces) |
Repeatability |
0.01? ( ±0.003? via interfaces) |
Frequency |
Range |
0.1Hz to 700 kHz |
Accuracy |
±0.1% (23? ± 5?) |
Inputs |
Voltage Range
Current ranges using Options 02F/B
All voltage and current input combinations are possible |
1mV rms to 350V rms (500V pk)
20mA - 2Arms range Opt. 02F/B 1R00 5Hz - 20kHz
1.0Arms to 20Arms range Opt. 02F/B 0R01 5Hz - 20kHz |
Bandwidth |
0.01Hz to 700kHz |
Isolation (inputs to ground) |
2000V |
Impedance |
2 Mohm and 10pF |
Environment |
Temperature |
Operating |
0?C to + 50?C |
Storage |
-40?C to +70?C |
Humidity (non-condensing) |
10% to 90% RH |
Dielectric Strength |
Inputs to case or AC line input |
2kV AC 50/60Hz 1 minute |
Power supply to case |
2kV AC 50/60Hz 1 minute |
Power Requirements |
Line input voltage |
110V or 220V AC ± 10% |
Line input frequency |
50/60Hz |
Line input fuse |
315mAT |
Line input conformance |
IEC348 Class I; IEC 1000-3-2;
IEC801 Parts 1 to 4; EN55011 |
Power consumption |
25VA |
Mechanical |
Weight |
5kg (6kg in delivery packaging) |
Dimensions (W x H x D) |
215mm x 144mm x 390mm |